FH-3000 Fully
Automatic
discrete Chemistry Analyzer

The independent
high-performance mixing needle design
Thoroughly part
of the solution easy blending condensate
sample issue

Integrated
intelligent sample needle, needle reagents
The latest
generation of liquid level sensors single
needle design, Dual completely eliminated
conductive liquid level sensor method of
cross-contamination and needles electrolytic
conductivity of the defects. With the volume
with tracking technology, needles and
samples, reagents contacts between at least
greatly reduce the impact of
cross-contamination.

High efficiency
color cup cleaning system

High precision
syringe(2set)
Samples and
reagents were used syringes independent high
precision, accuracy, faster, sample
Quantities of
stepping 0.1ul reached, the reagents can be
reached 1ul stepper.

Preset
detection project
Liver function
(ALT)
(AST)
(ALP)
(γ-GGT)
(TBA)
(TBIL)
(GPDA)
(TPRO)
(ALB)
(α-AFU)
(CHE)
(DBIL)
Renal function
(GLU)
(UREA)
(CREA)
(FMN)
(UA)
Blood Lipids
(CHOL)
(HDL-C)
(TRIG)
(LDL-C)
(APOA-1/B)
Myocardial
enzyme
(CK)
(CKMB)
(α-HBDH)
(LDH)
Pancreas
α-(AMY)
(LIP)
System models:
Automatic discrete
Experimental
models:Optional
type, the project team, volume, emergency
Testing speed:300T/H(Non-ion
electrode)
Determination:
destination, Kinetic,
turbidimetry, fixed-time law.
Samples / reagent needle:
An independent
samples and reagents acupuncture needles
with automated liquid level sensors, with
volume tracking
Cleaning:Disk
colorimetric reaction Cup(120) automatic
cleaning, recycling
Needle Cleaning:
sample and reagents needle acupuncture Both
internal and external wall washing, and
reduce the volume of pollution
Mixer:
Mixing independent needle
ensure full automatic reaction Stir
Reaction Cup:
120 of 6mm diameter thoroughly purple
Plexiglas’s Cup, Cup blank automatic
detection, repeated use
Samples Cup:May
also placed 60 original sample cup or tube
(expandable to 90)
Reagent:A
place for R1 R2 kit of a total of 45 at
once. (Expandable to 90)
Wavelength
Measurement:340、405、510、546、578、600、660、700..nm
(Single / dual-wavelength)
Calibration:
linear / non-linear multi-point calibration
Quality control:
can join the quality
control test any time, the same item can be
up to six species of quality control.
Programming projects:
Programming
can be arbitrary test
Light source:
12V halogen lamp
Absorbance:The
linear range 0.000-3.000Abs
Precision:0.0001
Abs
Reagents:
open (domestic, import
reagents may)
Reaction solution:
Minimum reaction
solution 180ul
Optical systems:
optical fiber after
spectrophotometry
Temperature control:
reaction temperature was 37ºC, reagents
fluorine-free refrigerated warehouses.
Sample volume:2-50ul
Reagents volune:R1
25-300ul,R2
10—150ul
Operating
system:Windows2000
or higher
Power:
220V 50HZ 1KVA
Dimensions:83cm
×118cm
×102cm
Weight: